Portable Ventilator (HV-2010)

Main Feature:
Gas driven electronically controlled.
Time switch, volume control, pressure limit ventilation mode.
I: E function
LCD display ventilation parameters, preset parameters alarm indication and breathing waveform
Applied for adult and child
Main Specification:
Power: AC 110~22OV10%, 50 Hz~60Hz1Hz, DC 12V 1.5A.
Tidal volume: 0~1.5L
Ventilation volume per minute: 0L~20L
Frequency: 4~99/min. (1~40/min under SIMV)
I/E Ratio: 3: 1~1: 5
Inhalation trigger pressure: -2.0~+2.0Kpa
O2 concentration: 48~100% adjustable.

Pressure limitation: 1~5.5KPa
Safety Pressure ≤ 6KPa
Ventilation monitoring:
Tidal volume, Minute ventilation, Frequency, Airway Pressure. Waveform of time-airway Pressure.
SPONT inhalation triggers indication, power indication.
Alarm: (Audible and visible reminding)
No tidal volume, ventilation volume is too high (low), airway Pressure is too high (low), frequency
Is too high, AC Power off, back up battery voltage is too low
Gas source: Oxygen or Purify air
Working Pressure: 0.4Mpa. Allowable range: 0.3~0.5MPa
O2 cylinder: 3L (Max Pressure 15MPa)
Internal back-up battery: DC 12V ≥ 3 hours.
Weight: ≤ 4Kg
Volume: 310mm x 190mm x 115mm (L x H x W).
Reserve Condition: Temperature: -10~40 c, Relative humidity: ≤ 90%
Working condition: Temperature: 5~40 c, . Relative humidity: ≤ 80%